Startup & the Setup Wizard
The Setup Wizard
3
-
8
LevelPRO User Manual
Thermo Scientific
6.
Click the Next button to proceed to the next screen.
Figure 3-5.
Wizard Type Selection
7.
Select Level Setup.
Figure 3-6.
Source Head, Isotope & Material Type Selection
8.
Designate the source head type, radioactive isotope and process material type by
clicking the dropdown selections next to each respective option.
Source Head Geometry – The source head geometry option allows the
program to include a very small correction factor in the single point
calibration calculation. The source head model number is stamped on the
identification plate on the source housing. If the model number of the source
head is unknown, or when using a head other than those on the dropdown
list, select the default, 5205.
Summary of Contents for LevelPRO
Page 2: ......
Page 4: ......
Page 6: ...This page intentionally left blank ...
Page 8: ...This page intentionally left blank ...
Page 16: ......
Page 102: ......
Page 127: ......