Troubleshooting
Investigating Peak Issues
110
ISQ User Guide
Thermo Scientific
Investigating Peak Issues
Table 3.
Troubleshooting Peak Issues in Your Data (Sheet 1 of 3)
Behavior
Characteristic
Cause
Remedy
Broadening General
Column
higher
than
optimum of column
Reduce the flow. Make sure vacuum
compensation is turned on.
Column flow lower than
optimum of column
Increase the flow.
Split flow is too low for
split injection
Increase the flow to 40-50 ml/min.
Performance of the column
has degraded
Test the column at the optimum flow
rate.
Injector is dirty
Clean or replace the liner.
Ion source is dirty
Clean the ion source and tune the
instrument.
Column is not far enough
into the transferline
The GC column does not extend into
the MS source. Use the column
measuring tool to confirm column
length. If the end of the column is inside
the transfer line, an excessive amount of
GC effluent will contact the inside wall.
Fronting
General
Column is overloaded
Decrease the injected amount and/or
analyte concentrations. Increase the split
ratio. Use a column with a thicker film.