3
Advanced Maintenance
Replacing Components of the Front Door
284
ISQ 7000 Hardware Manual
Thermo Scientific
Front Door Hinges and Supports
To replace the hinges and supports that connect the front door to the chassis
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the left side panel screw from the interior front
panel.
3. Slide the left side panel toward the back of the instrument and lift it off.
4. Look through the left side of the instrument and remove the nut and grounding washer.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
Figure 298.
Removing the Nut and Grounding Washer
Grounding Washer
Nut