Performing Routine Maintenance
Cleaning the Ion Source Cartridge
Thermo Scientific
ISQ 7000 Hardware Manual
73
2. Look for a small vertical
1
/
8
in. line engraved on the end of the barrel. There is a similar
engraved line on the sleeve of the ion source cartridge.
3. Position the line on the tool 5 mm to the left of the line on the ion source cartridge so
that the teeth of the locking ring will fit into the grooves on the end of the tool.
4. Push the ion source cartridge up into the barrel of the tool and twist it until the engraved
lines are aligned. The ion source cartridge should easily slide into place with very little
force.
Figure 67.
Aligning the Tool with the Ion Source Cartridge
Groove
on the
Sleeve
Groove
on End of
Tool
CAUTION INSTRUMENT DAMAGE:
If you push the ion source cartridge into the barrel
and it will not slide easily into position, make sure the components are assembled correctly
and have not slid out of place inside the sleeve. Using too much force when inserting the
ion source cartridge into the barrel can cause damage.