Initialize the sequencer
1.
In the instrument touchscreen main menu, tap
Initialize
.
Ion GeneStudio S5
The door, chip, and Reagent cartridge clamps unlock.
2.
When prompted, remove the Ion S5
™
ExT Wash Solution bottle to access the waste reservoir, then
remove and empty the waste reservoir.
3.
Reinstall the empty waste reservoir.
4.
Replace the expended Ion S5
™
ExT Sequencing Reagents cartridge with a new cartridge
equilibrated to room temperature.
5.
Ensure the new Ion S5
™
ExT Wash Solution bottle is thoroughly mixed. Then remove the red cap
and install.
1
2
3
6.
Ensure that the used sequencing chip from the previous run is properly seated in the chip clamp
and the chip clamp is pushed in all the way.
Chapter 5
Initialize the sequencer
Initialize the sequencer
5
Ion 520
™
& Ion 530
™
ExT Kit − Chef User Guide
47