6
Handling and Control
Setting the O2 Set Value
6-10
Heracell Vios 160i / 250i CR
Thermo Scientific
3. To set the CO
2
set value:
• Press the
Set value
key.
To increase the set value:
• Press the + key.
To reduce the set value:
• Press the - key.
4. To accept and save the set value:
• Press the
Enter
key.
• The system returns to the main menu. The CO
2
display shows the actual value currently measured
in the workspace.
Figure 6-10.
Setting the CO
2
Set Value
Setting the O
2
Set Value
This setting is possible only on versions with the optional O
2
/N
2
control.
1. Press the O
2
display key.
• The O
2
menu is displayed.
Note
Deactivating the CO
2
control loop:
To deactivate the CO
2
control, the set value is set to 0 %.
If the control loop is deactivated, error monitoring is enabled as well.
Note
Ventilate the inner chamber after changing the CO
2
setpoints so that no alarm occurs.
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...