5
Operation
Water Filling
5-4
Heracell Vios 160i / 250i CR
Thermo Scientific
5. Do not exceed the maximum level mark “MAX” (1/
). The maximum fill volume of the
water reservoir (5/
) is 3 l.
Figure 5-4.
“MAX” Fill Level Indicator
6. Wipe off excess water from the cover of the water reservoir.
7. Reinstall prefilter (2/
).
8. Reinstall the bottom shelf or close flap over cut-out for filling.
9. Make sure that the valves of the CO
2
/O
2
/N
2
supply system are open.
10. Turn on the device using the power switch.
11. Adjust set values for temperature and CO
2
/O
2
concentration on the iCan™ touchscreen.
Starting the Device:
12. Start the device with the auto-start routine (see
“Activating the Auto-Start Routine”
).
13. The auto-start progress indicator appears on the display, and the automatic start routine is begun.
14. The temperature control adjusts the temperature to the selected set value, humidity rises.
15. When temperature and relative humidity are constant, the automatic adjustment of the CO
2
/O
2
measuring system is performed.
16. The CO
2
/O
2
control supplies the set amount of CO
2
/O
2
.
17. If the auto-start routine has been completed, the progress indicator is hidden, and the main menu
is displayed. The device is now operational.
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...