4
Start-up
Gas Connection
4-14
Heracell Vios 160i / 250i CR
Thermo Scientific
The shelf should look like one of the following figures (for example 250 l version):
Installing the reinforced shelf
• Push the reinforced shelf onto the shelf supports and slightly raise the reinforced shelf so that the
pull-out stop can slide over the shelf supports.
Gas Connection
Rear side
Shelf guide
Pull-out stop
Front side
Note
Gas quality:
The gases must have one of the following qualities:
• Purity 99.5 % min
• medical gas quality.
CAUTION
Overpressure!
The operating pressure of the gas applied to the device must not exceed 1 bar (14.5 psi). If the gas is
supplied at a higher pressure, the valves integral to the device may not close correctly and the gas
supply control may be impaired.
Set the gas supply to a range between 0.8 bar (11.6 psi) min. and 1.0 bar (14.5 psi) max. and make
sure that this pressure setting cannot be changed!