3
Description of the Device
Workspace Components
Thermo Scientific
Heracell Vios 160i / 250i CR
3-17
The shelving system components, the airbox, the air duct and the cover of the water reservoir can be
removed without any need for tools, leaving only the easily treated, surface-reduced interior container
for cleaning and manual disinfection of the unit.
Glass Door and Optional Gas-Tight Screen
The standard version of the
Heracell Vios 160i CR
and
Heracell Vios 250i CR
units is equipped with a
single-piece door made of tempered safety glass.
Figure 3-9.
Heracell Vios 160i / 250i CR
with Glass Door
Due to the significantly smaller cross-section of the openings for accessing the samples, devices that are
equipped with the optional gas-tight screen have a considerably lower contamination hazard and the
recovery times of the incubation parameters are shorter for:
• workspace temperature,
• CO
2
concentration,
• O
2
/N
2
concentration,
• relative humidity.
Gas-tight screens are available as options for both types of products:
•
Heracell Vios 160i CR
: Threefold gas-tight screen or sixfold gas-tight screen – with or without Cell
Locker System (see Appendix) (see
),
•
Heracell Vios 250i CR
: Sixfold gas-tight screen.
Note
Oxidation of copper components:
When exposed to heat and humid air, the copper material of the interior container will oxidize.
This results in a discoloration of the copper components during the test run for checking the device
functions.
Do not remove the oxide layer during routine cleaning as the effect of the copper material is based
on it.
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...