3
Description of the Device
Heracell Vios 160i / 250i CR Back View
3-4
Heracell Vios 160i / 250i CR
Thermo Scientific
Heracell Vios 160i / 250i CR
Back View
Figure 3-2.
Heracell Vios 160i / 250i CR
Back View
1. Control box with supply interface for combined gas connection (optional) CO
2
and O
2
/N
2
2. Diagram: CO
2
and O
2
/N
2
gas connection without gas monitoring
3. O
2
/N
2
gas cylinder
4. CO
2
gas cylinder
Safety Devices
Heracell Vios 160i / 250i CR
The device is equipped with the following safety devices:
• A door switch interrupts the CO
2
/O
2
/N
2
supply and the workspace heating when the glass door is
opened.
• An overtemperature protection protects the samples from harmful overheating in case of failures.
• A pressure compensation opening ensures pressure compensation in the device workspace.
• The alarm relay circuit uses audible and visual alarms to indicate errors during operation.
• An externally mounted light provides a visual warning of any malfunction of the internal particle
extraction system.
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...