![Thermo Scientific Heracell Vios 160i CR Operating Instructions Manual Download Page 26](http://html1.mh-extra.com/html/thermo-scientific/heracell-vios-160i-cr/heracell-vios-160i-cr_operating-instructions-manual_1104558026.webp)
2
Installation of the Device
Site Ventilation
2-2
Heracell Vios 160i / 250i CR
Thermo Scientific
Site Ventilation
When CO
2
/O
2
/N
2
is supplied, the work space of the incubator is slightly pressurized. The pressure is
released through the pressure compensation opening into the operating room.
Pressure compensation and any opening of the glass door/gas-tight screen during the operation of the
device will release very small quantities of CO
2
/O
2
/N
2
into the operating room. The room ventilation
must be capable of carrying the released gas safely off into the open.
In addition, heat dissipating from the device during continuous operation may cause a change in the
room climate.
• Therefore, the
Heracell Vios 160i / 250i CR
must only be installed in rooms with sufficient
ventilation.
• Do not install the device in room recesses without ventilation.
• The room ventilation should be a technical ventilation that complies with the national
requirements for laboratories or some other suitable ventilation system with appropriate capacity.
Space Requirements
Figure 2-1.
Device Dimensions
WARNING
EMERGENCY STOP!
The mains socket must be accessible at all times for the case of an EMERGENCY STOP. The
mains socket must be easily identifiable by the user and freely accessible at all times.
Note
Accessibility of the devices:
To ensure the accessibility for care and maintenance works, keep larger side and rear distances. The
swivel-out for units with sixfold gas tight screen with Cell Locker must have a clearance to the wall
directly at the hinge side of >= 31 cm / 12.2”.
All dimensions in mm
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...