6
Handling and Control
Error Messages
Thermo Scientific
Heracell Vios 160i / 250i CR
6-45
Figure 6-52.
Failure Message Overtemperature
2. To exit the error display:
• Press the
End
key.
• The error message disappears.
• The temperature display field is displayed with a red bounding box.
3. Resetting the error message:
• Turn the device off.
4. Open the door and allow the workspace to cool down.
5. Turn the device on.
When overtemperature protection kicks back in although possible causes (see error log) have been
remedied, shut down the unit and call technical field service.
Measures After Power Outage
After a power outage, the formation of condensation can occur at the sensors due to moisture. This can
impair the sensor function and lead to erroneous values being displayed or to failure messages (sensor
breakage, see
).
The following measures must be taken to ensure safe operation:
1. Drain the water and wipe the interior dry.
2. Let the device warm up without water for 1 hour to 55 °C.
3. Then let the device cool down with the doors opened.
4. After this, start the device at incubation temperature according to Chapter
.
Alternatively, or when the above mentioned measures were not effective, start a sterilization run at
180 °C. Refer to Chapter
“Steri-Run Sterilization Routine”
The sterilization routine can be interrupted after approx. 1 hour. The sensors should then have dried.
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...