6
Handling and Control
Auto-Start Function
6-14
Heracell Vios 160i / 250i CR
Thermo Scientific
Activating the Auto-Start Routine
Preparations for the start:
1. Make sure that the valves of the gas supply system are open.
2. Fill the reservoir with purified water. The filling level must not exceed the upper filling level mark.
Launching the auto-start routine:
1. Press the auto-start key.
• The auto-start instruction menu is displayed.
Figure 6-13.
Activating the Auto-Start Routine
2. To exit from the auto-start instruction menu and stop auto-start:
• Press the
End
key.
3. Refill water, if necessary.
4. Activate the auto-start routine:
• Press the
Start
key.
5. To air the workspace, open both device doors. When the audible alarm sounds after 30 seconds.
Auto-start routine aborted due to error:
The auto-start routine is aborted if:
• an error is detected in the temperature control loop,
• an error is sensed in the CO
2
control circuit,
• the water filling level is insufficient,
• the CO
2
value to be set is out of tolerance.
Dry execution of auto-start routine:
If the routine is to be run dry (without water in the water reservoir of the workspace), the water level
sensor must be disabled prior to its start (chapter
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...