Setting up
Setting up
Thermo Scientific
19
Table 10.
Order of ECM electrodes from rod to probe
Order
Electrode
Color code
Color of
connecting wire
1
White Assembly A (not connected)
-
-
2
Reference electrode
Brown
Brown
3
pH electrode
White
White
4
EC Assembly
Transparent
Black
5
Spacer slice
Transparent
-
6
White Assembly B
Transparent
Transparent
Figure 16.
Connecting wires and electrodes
Note
Calibrate the tests and run quality controls after working with the ECM unit.