Thermo Scientific
G100 Range CO
2
Analyser
9-1
9
Service
The G100 instrument should be regularly serviced to ensure correct operation and accurate readings.
The manufacturer recommends a full service and recalibration
every 12 months
.
User Serviceable Parts
The following parts are supplied by your instrument manufacturer and can be user serviced:
Note
There are no user serviceable parts inside the instrument. Please do not attempt any
repair as this may invalidate any warranty supplied with your instrument.
Sample Filter
This should be regularly inspected for damage or discolou-
ration and changed if needed. The instrument should never
be operated without the sample filter as this may result in
water or dust entering the instrument. The filter should be
changed immediately if water can be seen. Failure to do so
can damage the instrument.
Sample Tubing
Always ensure that sample tubes are not contaminated or
damaged.
Summary of Contents for G100 Series
Page 1: ...Thermo Fisher Scientific G100 Range CO2 Analyser Operating Manual 50121623 12 8 09 ...
Page 6: ...Manual Guidelines Notes 1 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 34: ...Calibration Calibration Record 7 6 G100 Range CO2 Analyser Thermo Scientific ...
Page 40: ...Service 9 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 42: ...Warranty Policy 10 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 46: ...Event Log 12 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 48: ...Sample Certificate of Calibration 13 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 51: ...Thermo Scientific G100 Range CO2 Analyser 15 1 15 EC Declaration of Conformity ...
Page 52: ...EC Declaration of Conformity 15 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 57: ......
Page 58: ......
Page 59: ......
Page 60: ......