Problem Solving
Self-test Warning Messages
8-2
G100 Range CO
2
Analyser
Thermo Scientific
Self-test Warning Messages
The following warnings may be displayed during the self-test period when the instrument is switched
ON.
Warning
Description
CheckMemory
The instrument only has space to store less than 50 readings
before it is full. The exact number can be checked using the
‘View Readings’ option.
Memory Full
There is no more space in memory to store readings. Both
the store and log options will be disabled until the memory
is cleared. The readings should be downloaded to PC using
the optional download software before memory is cleared.
Battery Low
The instrument does not have enough power to operate for a
full day. The instrument should be recharged or connected
to an external power supply.
Service Due
It has been 12 months (or more) since your instrument was
returned to the manufacturer for a service. The performance
and accuracy of the instrument may be impaired.
Low Flow
The instrument’s gas inlet (or outlet) may be blocked. This
warning is most commonly caused by a water-logged or
dirty sample filter. Change the sample filter and check for
obvious blockages in the sample tubes. Alternatively, a
small amount of adjustment can be made to the low flow
detection point to compensate for minor changes in the per-
formance of the pump fitted to the instrument.
Check CO
2
Cal.
This warning is most commonly caused by an incorrect user
calibration. Try recalibrating the sensor or press ‘Return to
factory settings’. If the warning persists it may be caused by
dirt or damage to the infrared sensor. The instrument will
need to be returned to the manufacturer for service/repair.
Ref. Fault
This may be caused by dirt or damage to the infrared sensor
in the instrument. The instrument will need to be returned to
the manufacturer for service/repair.
*
Invalid
Config.
The instrument has detected a problem with the configura-
tion parameters. The instrument will need to be returned to
the manufacturer for service/repair. Most likely to be caused
after firmware update.
Change O
2
Cell
The oxygen cell has not been changed for at least 2 years; its
performance and accuracy may be impaired. The instrument
will need to be returned to the manufacturer for service.
Change Pump
The pump has exceeded its recommended run-time and
should be changed. The instrument will need to be returned
to the manufacturer for service.
Summary of Contents for G100 Series
Page 1: ...Thermo Fisher Scientific G100 Range CO2 Analyser Operating Manual 50121623 12 8 09 ...
Page 6: ...Manual Guidelines Notes 1 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 34: ...Calibration Calibration Record 7 6 G100 Range CO2 Analyser Thermo Scientific ...
Page 40: ...Service 9 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 42: ...Warranty Policy 10 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 46: ...Event Log 12 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 48: ...Sample Certificate of Calibration 13 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 51: ...Thermo Scientific G100 Range CO2 Analyser 15 1 15 EC Declaration of Conformity ...
Page 52: ...EC Declaration of Conformity 15 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 57: ......
Page 58: ......
Page 59: ......
Page 60: ......