Calibration
Calibration Record
Thermo Scientific
G100 Range CO
2
Analyser
7-5
Calibration Record
The G100 instruments have the facility to log user calibrations via the ‘Event Log’. This can be used as
an aid in ensuring that gas measurements are valid and accurate.
During calibration the instrument will record the following in the event log. For each entry the time
and date will be stored.
Event
Data recorded
Successful user zero CO
2
Type (N2 or Air) and Readings before and after
Successful user span CO
2
Target Value, Readings before and after
Failed user zero CO
2
Type (N2 or Air) and Reading
Failed user span CO
2
Target Value, Gas Reading
Return to factory settings
Note
If the calibration failed, then try again using a longer purge time or different target
gas. This event log can only be downloaded and viewed via the optional G100 Range -
Download Software. It cannot be viewed on the analyser screen.
Summary of Contents for G100 Series
Page 1: ...Thermo Fisher Scientific G100 Range CO2 Analyser Operating Manual 50121623 12 8 09 ...
Page 6: ...Manual Guidelines Notes 1 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 34: ...Calibration Calibration Record 7 6 G100 Range CO2 Analyser Thermo Scientific ...
Page 40: ...Service 9 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 42: ...Warranty Policy 10 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 46: ...Event Log 12 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 48: ...Sample Certificate of Calibration 13 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 51: ...Thermo Scientific G100 Range CO2 Analyser 15 1 15 EC Declaration of Conformity ...
Page 52: ...EC Declaration of Conformity 15 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 57: ......
Page 58: ......
Page 59: ......
Page 60: ......