Site requirements and safety information
Operating environment
Consider the following environmental factors when planning your work space.
Lifting and carrying
The FTIR Heated Valve Drawer weighs approximately 23 kgs (51 lbs). To avoid risk of injury, use
proper lifting techniques when lifting or moving the Heated Valve Drawer or other system
components.
CAUTION
Avoid risk of personal injury. Your Heated Valve Drawer cannot be lif-
ted safely by a single person. Lifting or moving the drawer requires two
people. Use proper lifting techniques when lifting or moving the drawer
or other system components.
Workspace considerations
The Heated Valve Drawer is measures approximately 48.26 cm wide, 30.5 cm high, and 38 cm deep
(19 in. wide, 12 in. high, and 15 in. deep). With the optional heated filter, the drawer is approximately
48.26 cm deep (19 in.)
The FTIR Heated Valve Drawer is designed to fit a standard open 19” rack
CAUTION
Do not position the FTIR Heated Valve Drawer so that it is difficult to
reach the cable connections and power switch on the rear panel.
Electrical service specification
CAUTION
Avoid shock hazard. Each wall outlet used must be equipped with a
ground. The ground must be a noncurrent-carrying wire connected to
earth ground at the main distribution box.
6
Modular Gas System User Guide
Thermo Scientific
Summary of Contents for FTIR
Page 1: ...Modular Gas System USERGUIDE 269 346401B July 2022 ...
Page 15: ... This page intentionally left blank ...
Page 38: ...using the pressure transducer Thermo Scientific Modular Gas System User Guide 33 Installation ...
Page 61: ... This page intentionally left blank ...
Page 79: ... This page intentionally left blank ...
Page 82: ...Eductor Panel Thermo Scientific Modular Gas System User Guide 77 Plumbing diagrams ...