Chapter 4
Oven Parameters
Column Oven
87
Oven Parameters
The
OVEN
menu contains the parameters for programming the oven temperature,
from an initial temperature to a final temperature, using up to three ramps during
the analytical run. It is possible to set a single (isothermal) or multiple ramp
program.
In Main Menu move the cursor until
Oven
is selected. Press
ENTER
to display the
OVEN
menu.
OVEN
Temp 100 100
Initial time 90.00
Ramp 1 120.0
Final temp 1 100
Final time 1 60.00
Ramp 2 Off
Table 4-1. Oven Parameters Menu
Parameter
Range
Description
OVEN
This line is the title bar
Temperature
On - Off
From 20 °C to 350 °C
in step of 1°C
Set On to display the actual and set point values. Set
the initial temperature in °C at which the oven is set
and conditioned before the GC enters the sampling
phase.
Initial Time
INF
From 0 to 99.99 minutes
Set the time in minutes. The oven remains at the
starting temperature after a programmed run has
begun. Setting INF, will cause the following menu
lines to disappear.
Ramp 1
On – Off – INF
From 0.1 to 120 °C/min
in step of 0.1 °C/min
Rate for the temperature ramp from Initial Temp to
Final temp 1. Set the rate increment to reach the
temperature set in Final temp 1. Set On to enable a
temperature ramp. Setting INF, the GC performs a
ballistic heating and the ramp duration is zero.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...