Using the Hydrogen Sensor
The lower limit of the hydrogen sensor is 0.5% in volume. You should adjust the
detection threshold to 1% in volume, which is 25% of the hydrogen lower limit of
explosion (4% in volume). In cases where the connections begin to leak or the
column breaks, the sensor alerts the operator. Then it automatically cuts off the
gas supply and heating to the active zones, and sweeps the column oven with
forced air ventilation. If the sensor detects anomalies or leaks during GC
operation due to instrument malfunction, the operator must immediately:
•
close the hydrogen supply
•
switch off the gas chromatograph
•
air out the room
The reliability of the sensor depends on careful maintenance. After the sensor is in
use, you must periodically check its operating performance and calibration as
recommended by the manufacturer. Refer to your hydrogen sensor’s instruction
manual for maintenance guidelines.
WARNING! Never use hydrogen in FOCUS GC unless your GC oven has a hydrogen sensor
installed.
Thermo Fisher Scientific CSEs are not authorized to install or repair any instrument
using hydrogen as a carrier gas unless the instrument is equipped with the appro-
priate sensor.
Cleaning and Decontamination
Normal usage of the FOCUS GC can cause the exterior to get dirty. Clean the
outer surfaces by wiping them with a cloth dampened with water.
In the event that a hazardous material is spilled on or in the instrument, clean the
spill according to the procedures in the Material Safety Data Sheet for that
substance.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...