Chapter 9
TCD Overview
Thermal Conductivity Detector (TCD)
173
Figure 9-1. TCD Filaments and Gas Flows
When the filaments are properly powered, they heat at a temperature (resistance)
that is a function of the thermal conductivity of the gas flowing through the
filaments. When a chromatographic component elutes in the analytical channel, a
change takes place in the heat transfer followed by a variation of the filament
temperature. The signal polarity is a function of the thermal conductivity of the
component relative to the reference gas and to the user-selected polarity of the
filament power supply.
WARNING! The TCD filaments are sensitive to impurities present in the carrier, reference, and
make-up gas supplies. To ensure correct detector operation, you should use oxygen
and water vapor traps in the carrier gas and the make-up gas supply lines. We
suggest that you install an OXICLEAR filter (PN 281 131 40) before connecting the
gas to the GC.
1.Measure Cell
A.Reference Gas
2.Reference Cell
B.Make-up Gas
3.Column
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...