Chapter 8
Flame Ionization Detector (FID)
FID Menu
164
FID Menu
The
DETECTOR
menu contains the detector control parameters.
In Main Menu, move the cursor until
Detector
is selected. Press
ENTER
to
display the
DETECTOR
menu.
DETECTOR (FID)
Flame Off
Temp 200 200
Signal pA (10.0)
Ing. Thresh. 2.0
Flameout retry Off
H2 On
Air On
Make-up Off
Table 8-1. FID Parameters
Parameter
Range
Description
DETECTOR
This line is the title bar
Flame
On –Off
This indicates the flame status: On, Off, Igniting,
Waiting or Out. Hydrogen and air flows are required to
light the flame.
Set On to turn on the flame flows. This happens only if
the Temp is 150 °C. If not, an error message is
displayed. The Out message is displayed when the flame
is inadvertently extinguished.
Set Off to turn off the flame flows.
Temp
100 - 375 °C
in steps of 1 °C Off
This indicates the detector base body temperature. Set 0
to turn off the heater and display the actual value.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...