Chapter 8
Flame Ionization Detector (FID)
FID Installation
162
FID Installation
This operation allows the correct installation of the FID on your FOCUS GC.
Material required
•
Jet for FID
•
Tool for jet
1. Place the jet into the detector base body housing and tighten it with the proper
tool. Ensure the jet is perfectly vertically aligned to avoid damaging its
ceramic part.
Figure 8-3. Jet for FID
2. Install the FID on the detector base body and secure it by using the fixing
screw on the front of the detector cell.
3. Carefully, connect the signal and ignition polarization cables coming from the
detector control card, to the detector cell.
1. Tool for jet
2. Jet
3. Detector base body
1
2
3
Summary of Contents for FOCUS GC