Chapter 8
FID Gas Supplies
Flame Ionization Detector (FID)
161
FID Gas Supplies
The performance of the FID is affected by the flow of the various gases through
the detector.
The fuel and make-up gases used for the FID are:
•
fuel gas: hydrogen and air
•
make-up gas: nitrogen (recommended) or helium
Detector gases are factory pre-set to:
•
350 mL/min for air
•
35 mL/min for hydrogen
•
30 mL/min for make-up gas
To ensure the correct regulation of detector gases a minimum line pressure of 4
bar (400 kPa, 60 psi) must be provided to the instrument.
NOTE
Usually the air flow is about ten times the hydrogen flow to keep the flame lit.
For high sensitivity applications, it is essential to exclude all traces of organic
contamination from the chromatographic system and/or detector gas lines. Such
contamination may create ghost peaks in the chromatogram or, more often, an
unstable baseline.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...