Chapter 7
Column
Keeping Leaks Under Control
154
OPERATING SEQUENCE
Performing an Automatic Leak Check
When you perform an automatic leak check of the FOCUS GC system, the GC
measures the column flow with a true mass flow sensor and compares it to a
calculated flow value obtained from the original column constant to see if the
numbers match. The instrument detects a gas leak if there is a significant
difference between the two values.
To start the leak check operate as follows:
1. In Main Menu move the cursor until
Carrier
is selected, then press
ENTER
.
2. Scroll to
Leak check
and press
ENTER
to open
Leak Check
menu.
3. To start column evaluation, scroll to
Run leak check
and press
ENTER
.
4. The system is automatically pressurized with carrier gas and sealed to
perform leak check.
NOTE
To exit
LEAK CHECK
menu without performing column evaluation, scroll to Exit w/o leak
check and press ENTER.
Leak check cannot be started in the following cases:
CARRIER (He)
Leak Check
LEAK CHECK
Run leak check
Exit w/o leak check
LEAK CHECK
Oven not ready
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...