
Chapter 7
Keeping Leaks Under Control
Column
147
OPERATING SEQUENCE
Manual Checking for Leaks
Before you begin this sequence, you must install the column into the injector, but
not into the detector base body.
Materials needed:
•
silicon rubber septum of any dimension
1. Carefully push the detector end of the capillary column into the rubber septum
to seal it.
2. Set split and purge valves Off. See
3. Increase the carrier gas pressure to 150–200 kPa and allow the column and
injector pressure to stabilize. This can take up to 30 seconds.
4. Reduce the pressure to 50 kPa.
5. Observe the actual pressure. In a leak-tight system, the pressure should not
drop more than 1 kPa/minute.
6. If your installed column is leak tight, remove the septum and prepare the end
of the column for installation into the detector.
7. If it is not leak tight, check the tightness of the column ferrule and repeat the
leak check sequence.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...