Chapter 7
Installation
Column
133
Preparing a Capillary Column
To prepare metal or fused silica capillary column operate as follows:
CAUTION
Wear safety glasses to protect your eyes from flying particles while handling,
cutting, or installing columns. Be careful handling columns to avoid accidental hand
injuries.
Fused Silica Capillary Column
To obtain a correct cut of fused silica capillary column, use ceramic scoring wafer
or sapphire scribe.
Materials required:
•
ceramic scoring wafer
1. Hold the capillary column between your thumb and index finger with the
column extending past the tip of your index finger.
2. Score the column very gently. Excessive force could crush the column end.
3. Snap off the end of the column.
4. Inspect the column end for an even, flat cut.
Metal Capillary Column
To obtain a correct cut of metal capillary column, use a triangular file.
Materials required:
•
Triangular file
1. Hold the capillary column between your thumb and index finger with the
column extending past the tip of your index finger.
2. Score all round the column very gently.
3. Snap off the end of the column.
4. Inspect the column end for an even, flat cut.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...