Chapter 5
Split/Splitless Injector (S/SL)
S/SL Inlet Parameters
110
OPERATING SEQUENCE
Programming the Split Mode
In split injection, only a portion of the sample transfers to the column. Most of it
discharges through the splitting line. The ratio between the split flow and the
column flow defines the amount of sample that enters the chromatographic
system. The split and column flows must be set to obtain the correct split ratio
necessary for the analysis.
Before you begin programming, do the following:
•
Verify that a column is correctly installed, the correct liner is in the injector,
and the system is free of leaks.
•
Program the carrier gas flow as described in Chapter 3.
Proceed as follows:
Use
INC
and
DEC
keys to increase or decrease the parameter values.
1. In Main Menu move the cursor until Inlet is selected. Press
ENTER
to display
the
INLET
menu.
2. In the Mode line select
Split
.
3. Scroll to
Temp
and set the appropriate value.
4. Specify the split flow or the split ratio. To set
Split Flow
enter the value in
the relevant line. The split ratio will be calculated for you. To set
Split
Ratio
, enter the value in the relevant line. The split flow will be calculated
for you.
5. If constant septum purge is required, scroll to
Constant Septum Purge
and set On to activate a constant septum purge. If constant septum purge is not
required, set the Constant Septum Purge Off. Scroll to
Stop Purge For:
and enter the time the purge flow should be interrupted.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...