Eutech DO 1730 Dissolved Oxygen Meter 27
4.1.5 DO Calibration Example
The following is an example of a DO calibration for 100%
percent saturation.
Press the “Cal” key to start the DO calibration. The
calibration type is shown.
Rinse the DO electrode and place into the prepared
calibration chamber.
Wait for the DO value to stabilize.
a.
When the reading is unstable, the stopwatch
icon is shown and the reading flashes.
b.
When the reading is stable, the checkmark icon
is shown and the reading is solid.
Once the reading is stable, press the “Enter” key.
The meter will save the calibration.
The calculated slope value is shown and then the
Calibration Done screen is shown.
The meter will proceed to the main measure mode.