4
Connecting the EASY-Spray Column Plumbing
Setting the Column Temperature
Thermo Scientific
EASY-Spray Series Ion Source User Guide
21
To install the EASY-Spray column
1. (Optional) Follow the procedure
To install the optional Tee holder under the column
2. Follow the procedure
To adjust the emitter tip position
.
3. Remove the protective cap on the end of the EASY-Spray column.
4. With the cable on the bottom, slowly push the tip of the EASY-Spray column into the
column holder opening until it clicks into place.
As the EASY-Spray column moves forward into the ion source, the protective cover over
the emitter retracts to expose the emitter tip.
5. Connect the cable to the front output heater socket.
6. Follow the appropriate procedure in
Examples of LC Connections to the EASY-Spray
Setting the Column Temperature
Before you begin, turn the temperature control knob to the lowest temperature setting.
To set the temperature for the EASY-Spray column
Do one of the following:
• Manually turn the ion source’s temperature control knob.
• (EASY-nLC 1200 instrument) Use either the Thermo Xcalibur Instrument Setup
window, or use the EASY-nLC 1200 instrument’s touch screen as follows:
i.
On the touch-screen’s Home page (top, right corner), press the
EASY-Spray
icon, .
Note
You can skip this step after the first time if you have not adjusted the column
holder’s position.
CAUTION High voltage.
To prevent electric shock, never insert a foreign object,
especially a conductive part, into the ion source opening. The opening is intended
only for the emitter positioning tool and the EASY-Spray column.
Note
Do not pull on the cable from the EASY-Spray column. Use the plug to connect
and disconnect the column.