6
Using EA IsoLink IRMS System for CNSOH
Starting a Sulfur Measurement
114
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
22. Fill in the correct masses (64 and 66 replace for example 44, 45 and 46) to the appropriate cups
specific for your IRMS.
When highlighting the specific gas configuration by a click on its row, the number of cups required
for measurement is displayed together with the assigned masses.
23. Select a
Calibration
, which is valid for the selected cups.
24. Press the
Save & Close
button
.
Starting a Sulfur Measurement
EA IsoLink Setting
See
Note
A value of -1 denotes unlimited
Table 24.
Setting for EA IsoLink
Carrier
180 mL/min
Oxygen
25 mL/min
Reference
20 mL/min
Cycle run time
1000 s
Sampling delay
10 s
Oxygen Injection End
3 s (or higher, depending on sample size and sample type)