Maintaining the DSQ II
Maintaining the Rotary-Vane Pump
68
DSQ II Hardware Manual
Thermo Scientific
Figure 39.
Rotary- Vane Pump Components
4. Disconnect the foreline vacuum hose.
5. Place the rotary-vane pump on a bench.
6. Drain the old oil.
a.
Remove one of the oil filler plugs.
b. Using gloves, remove the drain plug, and allow the oil to drain into a suitable
container.
c.
Dispose of the spent oil according to local environmental regulations.
d. Replace the drain plug.
1
Foreline Vacuum Hose
8
Drain Plug
2
Inlet Port
9
MIN (minimum level mark)
3
Gas-Ballast Control
10
Mode Selector
4
Oil Filler Plugs
11
On/Off Switch
5
Outlet Port
12
Handle
6
MAX (maximum level mark)
13
Voltage Indicator
7
Oil-Level Sight-glass
14
Electrical Inlet Connector
14
3
1
2
5
6
10
9
11
13
12
7
4
8
WARNING - LIFTING HAZARD.
Use a proper lifting technique with the rotary-vane
pump because it weighs approximately 22 kg (50 lbs).
WARNING - HAND AND CHEMICAL HAZARD.
Wear impermeable laboratory gloves
when changing oil.
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...