Maintaining the DSQ II
Cleaning the Ion Source Components
52
DSQ II Hardware Manual
Thermo Scientific
d. Attach the 3- and 8-pin connectors.
e.
Move the ion source left and right to ensure it is snapped in place. If it does not
move, it is in place. If the ion source is not in its groove properly, it could affect
performance and your ability to insert or remove the ion volume with an inlet valve.
f.
Remove the lens alignment tool.
g. Use the lens alignment tool to insert the ion volume. Make sure the arrow is facing up
and the ion volume is all the way in, which is typically when the flat surface of the ion
holder and heater board are exactly aligned. Any stainless steel that is outside the
board plane reduces ionization.
Cleaning the Prefilter
To clean the prefilter:
1. Remove the prefilter.
Figure 30.
Removing the Prefilter
1
Prefilter
4
Prefilter Clamp
2
Top Prefilter Connector
5
Bottom Prefilter Connector
3
Four Prefilter Clamp Screws
1
2
5
3
4