Vacuum System and Gas Inlets
Gas Inlet Components
Thermo Scientific
DSQ II Hardware Manual
163
Calibration Gas Flow Module
The calibration gas flow module controls the flow of calibration compound into the ion
source and is mounted at the back left of the DSQ II. The calibration compound provides
reference peaks to tune and calibrate the DSQ II.
The module can deliver two different flows of calibration compound. One is a low flow for EI
and NICI. The other is a higher flow used for PICI. The module permits calibration
compound to enter the ion source through 1/8 in. o.d. tubing. The module is factory charged
with 0.1 mL of perfluorotributylamine, which is adequate for approximately one year of
operation.
Figure 83.
Calibration Gas Flow Module
1
Gas Fittings
4
Two Screws
2
Foreline Adapter Tubing (not shown) 5
Two Twisted Pair Wires
3
Calibration Compound Vial
6
Calibration Gas Flow Module
(PN 119246-0001)
3
1
2
6
4
5