4
Vacuum System and Gas Inlets
This chapter describes the theory of operations for the vacuum system and gas inlets.
Vacuum System
Vacuum system components maintain the low pressure necessary for the ion source,
quadrupole, and ion detector to operate properly. The vacuum manifold, which houses the
ion source, quadrupole, and ion detector, is pumped (or evacuated) by the high vacuum
(turbomolecular) pump. All components, except the rotary-vane pump, are located around
the vacuum manifold.
The vacuum system consists of the following components:
• Fore pressure gauge
• High vacuum pump
• Ion gauge (upgrade)
• Rotary-vane pump
• Vacuum manifold
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...