Troubleshooting
Tuning Symptoms
Thermo Scientific
DSQ II Hardware Manual
149
Mass assignment is unstable.
Possible Solutions
• The RF and DC have not stabilized. Allow the instrument to scan for about 1 min.
• There is a problem with the RF or DC. See
Signal response is unstable or shows unexpected drop-out.
Possible Solutions
• Autosampler rinse vials have septum pieces in them, which can clog or change the
amount of sample in the syringe. Use aluminum foil instead of septa for the rinse vials.
Special Teflon caps are also available for the AI/AS 3000 Series and the Tri-Plus.
• There is a problem with the filament or lens control. See
.
• There is an air leak. Check for leaks. See
.
• There is a high vacuum problem. See
.
• There is a contamination problem. See
.
• Dust has collected in the electron multiplier or on the conversion dynode. Blow clean,
dry gas on the electron multiplier and conversion dynode.
Tuning Symptoms
Tuning symptoms may be the result of a incorrectly adjusted manual control parameter. You
should suspect a tuning problem when an automatic tune fails. Automatic tune performs
several functions and symptoms or error messages indicate different problems. A tuning
symptom can usually be found by running diagnostics in the
DSQ II Tune
software.
Automatic Tune reports “Cannot perform RF frequency calibration”.
Possible Solutions
There is a problem with the RF control system. See
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...