2
Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
Thermo Scientific
DSQ II Hardware Manual
99
a.
Disconnect the five cables from the RF module.
b. Disconnect the three cables from the DC driver.
5. Remove the RF module.
a.
Remove the twelve screws holding the RF coil cover in place.
b. Remove the RF coil cover.
Figure 58.
RF Module with RF Coil Cover Removed
c.
Use forceps to disconnect the four RF feedthrough connectors.
d. Remove the bottom three and loosen the top three screws holding the RF module in
place.
e.
Remove the two screws holding the RF module to the front of the chassis.
f.
Lift the RF module up and out of its slots.
1
Four RF Feedthrough
Connectors
3
Two Screws Around the RF
Module
2
Six Screws Around the RF
Module
1
2
3
2
2
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...