UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Operating Instructions
Page
81
5.
If you want to use column identification
Attach the column ID chip card to the column. Two chip cards are provided in the
accessories kit of the module. Wrap the ribbon around the column, pass the shank of the
rivet through a hole, and press down to unite the two parts.
Fig. 40: Attaching the column ID to the column
6.
Press the column into the bottom part of the clip and reinstall the ring.
Fig. 41: Installing a column in the clip
7.
Connect the capillaries to the column and establish the fluid connections between the
separation column and the column switching valve.
For information about how to connect the capillaries to the valve, see (
→
page 80). When
connecting the capillaries, observe the general information about how to connect and
route the capillaries (
→
8.
Activate column identification if applicable (
→
page 129).