UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Page
70
Operating Instructions
Setting the Solvent Type (ProFlow Flow Meter)
To set the solvent type for the solvent that you are using, follow these steps:
1.
In Chromeleon, open the
Commands
dialog box for the NC pump (
→
page 89).
2.
Select
%A_Solvent
(or
%B_Solvent
) and then select the preferred solvent from the
list.
Tip:
If you are using a solvent with up to 2 % additive, it is sufficient to select
the main solvent.
Solvent mixtures with higher percentages of a component require a solvent
calibration.
Fig. 31: Setting the solvent type (ProFlow flow meter only)
If the solvent type that you are using is not yet available in the list, perform a solvent
calibration for the solvent to be defined in the flow meter and added to the list in
Chromeleon. See section 5.5.3.1, page 114.
Summary of Contents for Dionex UltiMate 3000 Series
Page 2: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Operating Instructions ...
Page 8: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page vi Operating Instructions ...
Page 24: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 16 Operating Instructions ...
Page 62: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 54 Operating Instructions ...
Page 92: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 84 Operating Instructions ...
Page 246: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 238 Operating Instructions ...
Page 262: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 254 Operating Instructions ...