7.3 Removing Gas and Humidity from the Column Compartment
When a certain concentration of gas or humidity is reached inside the column compartment
(while the door of the column chamber is closed), the related sensor is activated and the
Status
LED on the front panel door changes to red.
1.
Find and eliminate the leak (wearing appropriate protective clothing).
2.
Beeping stops for the current alarm when you open the front panel door.
3.
Ventilate the interior of the column chamber thoroughly; only then, close the front
panel door.
4.
If the sensor is not activated again after the door has been closed, you can resume
operation.
Tip:
The
Status
LED for the column compartment remains red as long as the sensor
is exposed to gas or humidity. If a message appeared on the front panel display,
select
Clear
on the navigation bar to remove the message from the display.