ERR:SUPPRESSOR SHORT [124]
This error may be caused by a malfunction in the suppressor.
To troubleshoot:
•
After closing the error message, press the
AES/SRS
DEVICE
SELECT
button to restart the suppressor. If the error reappears,
replace the suppressor. Refer to the suppressor manual for
instructions.
•
ERR:OVER CURRENT CR-TC [130]
This error may be caused by a malfunction in the CR-TC.
To troubleshoot:
•
Replace the CR-TC (see Section 6.5).
•
ERR:OVER CURRENT EGC1 [140]
This error may be caused by a malfunction in the EGC controller
electronics.
To troubleshoot:
•
Call Thermo Fisher Scientific for assistance. The Dionex RFC
electronics components cannot be serviced by the user.
•
ERR:OVER VOLTAGE EGC1 [141]
This error appears if you press the
AES/SRS
DEVICE SELECT
button
and the Dionex RFC cannot establish a connection with the
suppressor.
To troubleshoot:
•
Check the suppressor cable connections.
•
If the error persists, replace the suppressor. Refer to the
suppressor manual for instructions. Suppressor manuals are
included on the Thermo Scientific Reference Library DVD
(P/N 053891).