![Thermo Scientific Dionex ICS-5000 Operator'S Manual Download Page 408](http://html.mh-extra.com/html/thermo-scientific/dionex-ics-5000/dionex-ics-5000_operators-manual_1105008408.webp)
Dionex ICS-5000 Operator’s Manual
392
Doc. 065342-06 12/11
12. Wait until the pump pressure has stabilized (30 to 60 seconds) and then
turn on the cell voltage.
13. Reapply the electrode potential. The baseline will drift for more than
1 hour as the cell re-equilibrates. Peak area values may require up to
12 hours to stabilize.
9.28.5 Replacing a pH-Ag/AgCl Reference Electrode
When to Replace a pH-Ag/AgCl Reference Electrode
Replace the pH-Ag/AgCl reference electrode if performance problems occur
that are not corrected by regenerating the electrode. Performance problems
can include no pH readouts, a shift in Ag/AgCl reference potential or
incorrect readouts, baseline spikes, or a decreased response even with a
freshly polished working electrode. The pH-Ag/AgCl reference electrode
typically lasts from 3 months to 1 year, depending on use.
NOTE
To regenerate a pH-Ag/AgCl reference electrode, soak
it in a solution of 1 M KCl and 1 M HCl.
Items Needed
•
pH-Ag/AgCl reference electrode (P/N 061879)
pH-Ag/AgCl Reference Electrode Replacement Procedure
1. Follow the instructions in
to turn off the cell voltage, stop
the pump flow, and disconnect the cell from the detector.
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...