Gas and Humidity Sensors
The TC contains two sensors to detect any gas or humidity that may
accumulate inside the TC. When a certain concentration of gas or humidity is
reached (while the door is closed), the following events occur:
•
The corresponding sensor is activated.
•
The
ALARM
LED on the front of the TC lights (red), and remains red as
long as the sensor is exposed to gases or humidity.
•
The error is logged in the Chromeleon Audit Trail.
•
A beep sounds.
What to do when an alarm occurs:
1. Find and eliminate the leak.
2. If you are using Chromeleon 7: Open the Commands dialog box (press
the
F8
key). Select the TC and click the
Commands
tab. Click the
ClearAlarmLED
button.
3. If you are using Chromeleon 6.8: On the TC Control panel, click the
Clear Alarm
button.
4. Ventilate the interior of the TC thoroughly before closing the door and
resuming operation.
Adjusting the Sensor Sensitivity
Follow these steps to change the sensitivity setting of the gas or humidity
sensor, or to turn off a sensor:
1. Open the Chromeleon 7 ePanel Set or the Chromeleon 6.8 panel tabset.
2. Press the
F8
key to open the Commands dialog box.
3. Select the TC.
4. If you are using Chromeleon 7: Click the
Parameters
tab page for the
TC, click the
GasLeakSensor
or
HumidityLeakSensor
parameter.
Select the new sensitivity setting (or select
Off
to disable the sensor).
5. If you are using Chromeleon 6.8, click the plus sign beside
TC
to display
the items underneath. Under
TC_Temp
, select the
GasLeakSensor
or
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...