4 • System Plumbing
Doc. 065447-02 12/14
75
4.1.4
Connecting the Pump Outlet
If you did not install the Dionex ASTC 500 Anion Trap Column (see
), connect the pump outlet tubing directly to the pump pulse
damper.
On the pre-assembled tubing chain (see
), connect the black PEEK
tubing labeled
TO PUMP/DAMPER
to the pump pulse damper outlet. Keep the
remaining tubing in the chain connected.
4.1.5
Filling and Flushing the CD Regenerant Path
NOTE
If you are installing a system with an ED, go on to
and Flushing the ED Liquid Path” on page 78
.
If you are installing a system with a CD, follow the steps below to fill and
flush the suppressor and CRD regenerant pathways, including the connecting
path in the interior of the Dionex ICS-5000
+
IC Cube.
1. If you have not already done so, install the CES cartridge and CRD
cartridge (if used) in the Dionex ICS-5000
+
IC Cube.
2. In the Dionex ICS-5000
+
IC Cube Tubing Kit (P/N 072186), locate the
following two tubing assemblies: 65 mm (2.56 in) precision cut
(P/N 072188) and 125 mm (4.92 in) precision cut, labeled
VALVE PORT 3
(P/N 072189).
Capillary
IC
Figure 4-2. Capillary Pump Outlet Connection
Pulse
Damper
Outlet
Pump Outlet
Tubing
Capillary
IC