3 • Operation and Maintenance
Doc. 031857-03 4/06
3-7
3.4.2
Fill the Reservoir
Fill the reservoir with ASTM filtered, Type I (18-megohm) deionized
water.
3.4.3
Set the Eluent Level
After filling the reservoir, enter the volume of liquid in the reservoir in
one of the following locations:
•
On the
HOME
page in the
Eluent Level
field (see
•
In the
Eluent Bottle
field on the Chromeleon
Control panel (see
The ICS-2000 determines the eluent usage by monitoring the flow rate
and the length of time the pump is on. The ICS-2000 updates the field as
the eluent is used up. A warning first appears if the level falls below
200 mL. Warnings are repeated at 100 mL and 0 mL.
In order for the eluent level displayed on the screen to be accurate,
the level must be entered by the user when the reservoir is filled. The
ICS-2000 does not automatically detect when the reservoir is filled,
nor when it is empty.
Figure 3-7. Home Page: Setting the Eluent Level
Figure 3-8. Chromeleon Control Panel: Setting the Eluent Level
Enter the volume of
liquid in the reservoir
Note: The Eluent Level
field is updated as the
liquid is used up.
Enter the volume of
liquid in the reservoir
Note: The Eluent Bottle
field and slider gauge
are updated as the
liquid is used up.
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...