ICS-2000 Ion Chromatography System
B-114
Doc. 031857-03 4/06
B.21.1 Connect the Gas Source
1. Locate the barbed fitting (P/N 030071), pipe thread reducer
(P/N 030087), 3-mm (1/8-in) ID tubing (P/N 040793), and ICS-2000
air regulator accessory (P/N 060054). These parts must be ordered
separately.
2. Use the barbed fitting and pipe thread reducer to connect the 3-mm
(1/8-in) ID tubing to a clean gas source, regulated to between 0.55
and 0.83 MPa (80 and 120 psi).
3. Push the other end of the tubing onto the inlet of the air regulator (see
4. Remove the eluent reservoir cap.
5. Connect the gas line from the eluent reservoir cap to the gas line on
the air regulator outlet.s
6. Slip the air regulator accessory bracket over the neck of the eluent
Figure B-112. ICS-2000 Pressurized Eluent Reservoir Connections (Optional)
Gas Inlet Line
3-mm (1/8-in) ID
Connect to
gas source
Eluent Gas Line
Air Regulator
Accessory
Bracket