Setting Up the Eluent Generator
Doc. 031857-03 4/06
B-61
13. Remove the plug from
the
EGC-1
connector on
top of the ICS-2000.
14. Orient the cartridge’s
blue 4-pin electrical
cable connector as
shown in
.
Push it onto the
EGC-1
connector on the ICS-
2000. Twist the ring on
the cable connector
fingertight to secure it.
Figure B-58. Connecting the EluGen Cartridge Cable
Connecting the EluGen Cartridge Vent Line
1. Locate the EluGen Cartridge Vent Line Tubing Kit supplied with the
EluGen cartridge.
2. Slide the label onto one end of the tubing and then push the barbed
luer lock onto the tubing.
3. Unscrew and remove
the plug from the vent
opening on the side of
the EluGen cartridge
(see
Figure B-59. EluGen Cartridge Vent Plug
Vent Plug
(remove)