Setting Up the Eluent Generator
Doc. 031857-03 4/06
B-35
B.14 Setting Up the Eluent Generator
Before Beginning
Remove the EluGen cartridge (EGC) and the other items to be installed from their
shipping containers. Depending on the type of eluent to be generated, the other
items may include a Continuously Regenerated Trap Column (CR-TC), an EPM
Electrolytic pH Modifier, and/or an EGC CO
3
Mixer.
NOTE The EGC degas assembly is installed inside the EGC
holder at the factory.
Refer to the EluGen cartridge, CR-TC, and EPM manuals for the optimal
operating conditions for your application. The manuals are provided on the
Dionex Reference Library CD-ROM (P/N 053891).
The EluGen cartridge contains one of the following: a corrosive base
(KOH, LiOH, or NaOH), a corrosive acid (MSA), or a concentrated
K
2
CO
3
solution. Wear protective eyewear and gloves when handling
the cartridge.
La cartouche d'EluGen contient un de ce qui suit: une base corrosive
(KOH, LiOH, ou NaOH), un acide corrosif (MSA), ou une solution
concentrée de K
2
CO
3
. Porter des lunettes et des gants protectives en
manipulant la cartouche.
Die EluGen-Kassette enthält eine korrodierende Base (KOH, LiOH
oder NaOH), eine korrodierende Säure (MSA) oder eine konzentrierte
K
2
CO
3
-Lösung. Tragen Sie daher beim Umgang mit Kassette eine
Schutzbrille und Handschuhe.
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...