Dionex Easion Operator’s Manual
62
Doc. 155028-01 08/20
5.1.1
Wellness Panel Features
The Wellness panel provides the following features.
System Status
•
When the Dionex Easion is connected to Chromeleon, the
green
Connected
LED is displayed. When there is no
connection, the red
Disconnected
LED is displayed and all
other Wellness panel controls are disabled.
•
Click the pump
On
or
Off
button to turn the pump on or off,
respectively.
•
Click the
Log Serial
button to record the Dionex Easion
serial number in the Chromeleon audit trail.
•
The
Calibration
or
Diagnostic
box is green while a
calibration or diagnostic procedure is running.
•
The current electrochemical detector (
ECD
) conductivity,
pump
Flow
rate, and total system
Pressure
are displayed.
View and/or
Reset
Calibration
Parameters
•
Click the
Calibration Values
button to open a window that
displays the current calibration values, the previous values,
and the factory-set values.
•
You can reset the current calibration values to the previous or
factory-set values from the window.
Calibration
Pressure Transducer
•
Use the
Calibrate Offset
and
Calibrate Slope
buttons when
performing the pressure transducer calibration procedure (see
•
The
Offset
and
Slope
values and the date of the last
calibration are displayed.
Pump Flow Rate Calibration
•
Click the
Calibration Procedure
button to start the pump
flow rate calibration procedure (see
•
Click the
Log
button to record the calibration value in the
Chromeleon audit trail.