5 • Service
Doc. 065330-05 03/2020
103
5.10 Calibrating a Sample Pump
Do not calibrate a sample pump
unless
the volume removed from the sample
container does not match the sample volume requested in the
Load Sample
step
of your method.
Parts and Tools Needed
•
Allen wrench, 3/32-in
•
Sample container with approximately 120 mL of water/pump being calibrated
•
Balance
•
1.6-mm (1/16-in) ID tubing (P/N 070358) provided in the Ship Kit (standard
Dionex AutoTrace 280, P/N 071383; Dionex AutoTrace 280 PFAS,
P/N 22136-62005)
Calibrate the Pump
Perform the following procedure on each sample pump that requires calibration.
1. Run a single-step
Clean Sample Path
deionized water.
2. Select the Calibrate Pump diagnostic method:
a.
Press
LOAD
to display the
LOADING METHOD
screen.
b. Press
LOAD
to increment the displayed method to 31.
c.
Press
CONT
once to select method 31.
3. Press
CONT
to load the method.
4. Follow the on-screen instructions as they appear.
5. As instructed, weigh the collection vessel before and after pumping.
Determine the percentage difference between the
expected
volume (100 mL)
and the
delivered
volume.
For example, if the method pumps 103 mL instead of 100 mL, the volume is
off by 3%.
6. Locate the pump’s calibration screw in the hole on the right side of the pump